Abstract
A major concern in pediatric dentistry is maximizing risk management through optimal monitoring of respiratory function during sedation techniques, and Pulse Oximeter is one of the most popular devices for these purpose. Pulse Oximeter is non-invasive device for detecting the sensitive fraction between the saturated & desaturated hemoglobin. Several Studies proved that there is no significant difference between the SpO2 and SaO2. In this article, We examined three Pulse Oximeter (3300MX®,OX90®,BPM200®. The Pulse Oximeter using shorter pulse beat averaging showed more sensitive reaction to the statue of the patient and lower saturation data. We compared pair of Pulse Oximeter applied to one patient at one time. and BPM200® (using shorter pulse beat averaging) showed statistically low saturation.
Key Words: Average heart beat, Oxygen Saturation, Pulse Oximeter