POLYMERIZATION SHRINKAGE OF COMPOSITE RESIN USING DOUBLE CURING UNIT SYSTEM
단국대학교 치과대학 소아치과학교실 
Double curing unit system을 이용한 복합 레진의 광중합 수축에 관한 연구
한미란, 김종수, 유승훈
단국대학교 치과대학 소아치과학교실
Correspondence:  , 
Received: 12 September 2008   • Accepted: 17 February 2009
Abstract
As a part of an effort to minimize the polymerization shrinkage which is considered to be a major cause of failed bonds to tooth, newly designed 'Double LED system' was tested in the present study. Analyses were performed on the pattern of micro-leakage and the changes of strain which have occurred during the polymerization process. The results can be summarized as follows: 1. In the strain change, dramatic increase was observed with initiation of polymerization which was followed by subsequent gradual decrease with elapse of time in both the single LED system and double LED system. 2. The single LED system were shown to develop and maintain the maximum stress more than double LED system(p<0.05). 3. Less micro-leakage was found in the double LED system than in the single LED system(p<0.05). From the above-mentioned results, the double LED system can be a very useful tool in a sense of reducing polymerization shrinkage when compared to the single LED system. However, practical problems such as size of curing unit and its application method with its light intensity should be solved before its clinical application.
Key Words: Composite resin, Light curing, Polymerization shrinkage, Strain gauge


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
Seoul National University, Dental Hospital, B1-166 101, Daehak-ro, Jongno-gu, Seoul 03080, Republic of Korea
Tel: +82-70-4145-8875    Fax: +82-2-745-8875    E-mail: info@kapd.org                

Copyright © 2024 by Korean Academy of Pediatric Dentistry.

Developed in M2PI

Close layer
prev next